Loading...

Fume hood CompactLine DCL 1800 SS


Technical data

TECHNICAL DATA - DIMENSIONS
overall dims [mm]
width A1880
height B2325-2600
depth C960
working space dims (mm)
width D1750
window width D11565
height E1220
depth F635
TECHNICAL DATA
recommended airflow [m3/h]900 - 1500
required capacity of the ventilation system [m/s]0,3 - 0,5
nominal power [W]82
power supply230 VAC 50Hz
diameter of the air outlet stub [mm]Ø 200
water supplyG 1/2 '' male thread
sewage connection diameter [mm]Ø 50
electrical insulationclass I
standard ventilation systemdouble rear wall
illumination installed outside the working chamberLED, class A ++, through insulating glass / from operator panel
monitoring set standard:
- FM 550-A-0-E controller with a speed sensor AFS 100 in the fume hood window and the FA-0025-3 operator panel
option:
- ICM 500-F-EU controller with flow control
- ICM 500-FP-EU controller with flow control and cooperation with BMS
- other Schneider controllers available on request
raising the window standard:
- manual with counterweight (lock at a height of 500mm, maximum opening 810mm)
option:
- SC 500 - Auto Protect system - with foot switch
- SC 500 - Auto Protect system - with panel switch
frame and housing materialThe frame is made entirely of galvanized steel.
External housing made of galvanized and painted
with epoxy paint steel
working chamber wall finishinggalvanized steel painted with epoxy paint
worktopMonolithic ceramics with marine edge
worktop OPTION (available materials)- Phenolic resin, epoxy
- Stainless steel AISI 304, 316
underbench cabinet OPTION (different types)- metal, ventilated cabinet painted with epoxy paint
- cabinet for acids and alkalis made of foamed PVC
- explosion-proof cabinets Asecos 30 and 90 minutes
warranty24 months
manufacturerPOL-EKO-APARATURA

Download

Compact Line DCL-1800 SS

ikonka pdf Data Logger RT2014 EN