Fume Hood CompactLine DCL 1200 CR

TECHNICAL DATA - DIMENSIONS
overall dims [mm]
width A1280
height B2325-2600
depth C960
working space dims (mm)
width D1150
window width D1965
height E1220
depth F635
TECHNICAL DATA
recommended airflow [m3/h]600 - 950
required capacity of the ventilation system [m/s]0,3 - 0,5
nominal power [W]46
voltage**230 VAC 50Hz
diameter of the air outlet stub [mm]Ø 200
water supplyG 1/2 '' male thread
sewage connection diameter [mm]Ø 50
electrical insulationclass I
standard ventilation systemdouble rear wall
illumination installed outside the working chamberLED, class A ++, through insulating glass / from operator panel
monitoring set standard:
- FM 550-A-0-E controller with a speed sensor AFS 100 in the fume hood window and the FA-0025-3 operator panel
option:
- ICM 500-F-EU controller with flow control
- ICM 500-FP-EU controller with flow control and cooperation with BMS
- other Schneider controllers available on request
raising the window standard:
- manual with counterweight (lock at a height of 500mm, maximum opening 810mm)
option:
- SC 500 - Auto Protect system - with foot switch
- SC 500 - Auto Protect system - with panel switch
frame and housing materialThe frame is made entirely of galvanized steel.
External housing made of galvanized and painted
with epoxy paint steel
working chamber wall finishingBuchtal large-size ceramic
worktopMonolithic ceramics with marine edge
worktop OPTION (available materials)- Phenolic resin, epoxy
- Stainless steel AISI 304, 316
underbench cabinet OPTION (different types)- metal, ventilated cabinet painted with epoxy paint
- cabinet for acids and alkalis made of foamed PVC
- explosion-proof cabinets Asecos 30 and 90 minutes
warranty24 months
manufacturerPOL-EKO

Additional electric socket (400V)

Order number: D-S-EL400

Additional electric socket (400V)
Residual current device

Order number: D-WRP

Residual current device
Emergency stop button

Order number: D-GRZ

Emergency stop button
Main circuit breaker

Order number: D-MAIN

Main circuit breaker
EX Lamp

Order number: D-L-EX-XX.XX

EX Lamp
Additional water instalation

Order number: D-S-WA

Additional water instalation
Additional LPG instalation

Order number: D-S-GA-XXX

Additional LPG instalation
Advanced fume hood controller ICM500-F-EU

Order number: ICM500-F-EU-XX.XX

Advanced fume hood controller ICM500-F-EU
Advanced fume hood controller ICM500-FP-EU

Order number: ICM500-FP-EU-XX.XX

Advanced fume hood controller ICM500-FP-EU
Underbench cabinet (steel)

Order number: D-SZ-SS-Z-XX.XX

Underbench cabinet (steel)
Underbench cabinet (polypropylene)

Order number: D-SZ-PP-K-Z-XX.XX

Underbench cabinet (polypropylene)
Underbench cabinet (fireproof)

Order number: D-SZ-A90.50-XX.XX

Underbench cabinet (fireproof)
Automatic sash window

Order number: SC500

Automatic sash window
Side walls glazing

Order number: D-S-P

Side walls glazing
Scaffold

Order number: D-KRX-XX.XX

Scaffold
Cuvette for cabinets

Order number: D-SZ-KD-XX.XX

Cuvette for cabinets
Additional sink

Order number: D-SINK-NS

Additional sink
Non-standard worktop

Order number: D-WT-NS

Non-standard worktop
Funnel for gathering leachate

Order number: D-S.C.A.T.

Funnel for gathering leachate

Compact Line DCL-1200 CR