TECHNICAL DATA - DIMENSIONS | |
overall dims [mm] | |
width A | 1800 |
height B | 2385-2850 |
depth C | 1200 |
working space dims (mm) | |
width D | 1470 |
window width D1 | 1410 |
height E | 2145 |
depth F | 845 |
![]() | |
TECHNICAL DATA | |
recommended airflow [m3/h] | 900 - 1500 |
required capacity of the ventilation system [m/s] | 0,3 - 0,5 |
nominal power [W] | 82 |
voltage | 230 VAC 50Hz |
diameter of the air outlet stub [mm] | Ø 250 |
electrical insulation | class I |
standard ventilation system | double rear wall |
monitoring set | standard: - FM 550-A-0-E controller with a speed sensor AFS 100 in the fume hood window and the FA-0025-3 operator panel option: - ICM 500-F-EU controller with flow control - ICM 500-FP-EU controller with flow control and cooperation with BMS - other Schneider controllers available on request |
raising the window | standard: - manual with counterweight (lock at a height of 500mm, maximum opening 1850mm) option: - SC 500 - Auto Protect system - with foot switch - SC 500 - Auto Protect system - with panel switch |
frame and housing material | The frame is made entirely of galvanized steel. External housing made of galvanized and painted with epoxy paint steel |
working chamber wall finishing | galvanized steel painted with epoxy paint |
warranty | 24 months |
manufacturer | POL-EKO® |
img_photo | dygestorium-fume-hood-DCL-walk-in-1800 |
txt_opis_foto | The photo above is for reference only, may show additional options not included in standard equipment. The real appearance, particularly color and structure of the material may differ from the ones presented in the photo. |