TECHNICAL DATA | |
air convection | forced |
chamber capacity [l] | 245 |
working capacity [l] | 245 |
controller | microprocessor PID |
display | 7" full colour touch screen |
TEMPERATURE AND HUMIDITY | |
Working temperature range without humidity [°C] | 0…+70 (max 20°C below ambient temp.) |
Temperature variation (spatial) at 40°C [+/-°C] | ±0,3 |
Temperature fluctuation (time) at 40°C [+/-°C] | ±0,1 |
Working temperature range with humidity [°C] | +5…+70 (max 20°C below ambient temp.) |
temperature resolution every … [°C] | 0,1 |
Temperature variation (spatial) at 40°C, 75% RH [°C] | ±0,3 |
Temperature fluctuation (time) at 40°C, 75% RH [°C] | ±0,1 |
Temperature variation (spatial) at 25°C, 60% RH [°C] | ±0,2 |
Temperature fluctuation (time) at 25°C, 60% RH [°C] | ±0,1 |
temperature protection | class 3.3 to DIN 12880 |
relative humidity range [%] | 10…90 (see chart for details) |
humidity resolution every … [%] | 0,1 |
Humidity variation (spatial) at 40°C, 75% RH [+/-%RH] | < ±1,0 |
Humidity fluctuation (time) at 40°C, 75% RH [+/-%RH] | ±0,3 |
Humidity variation (spatial) at 25°C, 60% RH [+/-%RH] | ±0,8 |
Humidity fluctuation (time) at 25°C, 60% RH [+/-%RH] | ±0,2 |
Recovery time to set values after 30 sec door open at 40°C, 75% RH [min] | 10 |
CHAMBER | |
door type | 1 external solid door, 1 internal glass door |
interior | |
Smart PRO | acid-proof stainless steel to DIN 1.4301 |
IG Smart PRO | acid-proof stainless steel to DIN 1.4301 |
housing | |
Smart PRO | powder coated sheet |
IG Smart PRO | stainless steel linen finish |
overall dims [mm] /1/ | |
width A | 960 |
height B | 1140 |
depth C | 840 |
internal dims [mm] | |
width D | 600 |
height E | 800 |
depth F | 510 |
shelves (standard | max) | 3 | 10 |
max shelf workload [kg] /2/ | 25 |
max unit workload [kg] | 90 |
External water tank capacity [l] | 6 |
weight [kg] | 117 |
ELECTRICAL PARAMETERS | |
voltage | 230V 50-60Hz |
nominal power [W] | 2300 |
warranty | 24 months |
manufacturer | POL-EKO® |
all the above technical data refer to standard units (without optional accessories)
* - variation (K) calculated for chamber as: K= +/- (T average max. - T average min.) / 2
** - other power supplies on request
1 - width doesn’t include shelve for water tank - 140mm
2 - on uniformly loaded surface